Stage Top Incubator – Water Jacket Incubator

T range: from 10–15 °C below ambient (model dependent) up to 60 °C

Water Jacket IncubatorOkolab’s Water Jacket Incubating Chamber provides rock-solid incubation stability both above and below room temperature, thanks to the thermal properties of water. 

Water is pumped from an external bath into the chamber’s base and lid. The actual sample temperature is monitored inside the chamber, using the provided fine gauge thermocouple. The temperature controller uses this thermal feedback from the sample to update the water bath’s set point, ensuring a specimen temperature accuracy of ± 0.1°C.

Features
  • Temperature Range: from 10–15 °C below ambient (model dependent) up to 60 °C
  • Stable Sample Temperature: ± 0.1 °C accuracy using Sample Feedback Mode
  • Chamber Feedback Mode Option: ± 0.3 °C when room temperature remains stable; no temperature probe required
  • Water Circulation: occurs in both base and lid to minimize gradients
  • Modular Inserts: magnetic inserts support petri dishes, chamber slides, and multi-well plates
  • Perfusion Ports: built-in tubing inlets/outlets for microfluidics and perfusion applications

The Water Jacket Incubator provides superior incubation stability above and below room temperature by circulating water inside the base and lid of the water jacket chamber. Feedback is provided by sample temperature or by chamber temperature. Selection is done via OKO-TOUCH.

Okolab offers two Temperature controllers, according to the temperature range of interest.

H101-BASIC-BL – T range: [3°C above ambient; 60°C]

 

  • Temperature Range: from 3 above ambient to 60°C
  • Can operate in sample or chamber feedback mode
  • Temperature Accuracy on sample:
    • ±0.1°C in sample feedback mode
    • ±0.3°C in camber feedback mode, if room T remains within
  • Touch Screen Interface – OKO TOUCH
  • Compatible with:
    • Digital Gas Controllers: accuracy ±0.1%
    • Manual gas mixers: accuracy ±1%
    • Bubbling Humidification Unit: efficiency 75%
    • Active Humidity Controller: efficiency 95%
    • Water Jacket Objective Heater
    • Data Logger / Remote Operator SMART BOX

H101-CRYO-BL – T range: [5°C; 60°C]

  • Temperature Range: from 10/15°C BELOW Room temperature (depending on chamber model) to 60°C
  • Can operate in sample or chamber feedback mode
  • Temperature Accuracy on sample:
    • ±0.1°C in sample feedback mode
    • ±0.3°C in camber feedback mode, if room T remains within
  • Touch Screen Interface – OKO TOUCH
  • Compatible with:
    • Digital Gas Controllers: accuracy ±0.1%
    • Manual gas mixers: accuracy ±1%
    • Bubbling Humidification Unit: efficiency 75%
    • Active Humidity Controller: efficiency 95%
    • Water Jacket Objective Heater
    • Data Logger / Remote Operator SMART BOX

 

 

Gas Control: CO2 and O2 Precision

Digital Bold Line Gas Mixers connect directly to temperature controllers and are managed through the same user interface (OKO‑TOUCH).

Features

  • CO2 and O2 Accuracy: ± 0.1 %
  • Adjustable Output Flow: programmable via touch interface
  • Programmable Gas Cycles: for complex protocols
  • Optional Active Humidity Control: compatible with Bold Line systems
  • Data Logging & Software Integration: USB, SMART BOX, and third‑party microscopy platforms
  • Optional Compressed Air using Okolab air pump
  • PTFE Filter: removes particulates from gas streams

Humidity Module

Gas is humidified through a bubbling column immersed in the water bath, with a pre‑heating coil to prevent condensation and ensure thermal equilibrium before entering the incubation chamber.

  • Can be used in cooling and heating modes
  • Open-loop, 80 % relative humidity to the sample

The OKO‑TOUCH touchscreen provides centralized control of all connected Bold Line controllers.

It automatically detects system components and displays supported functions, including:

Features:

  • System power control
  • Set point definition (temperature and gas)
  • Flow rate adjustment
  • Temperature and gas cycle programming
  • Feedback mode selection (sample vs chamber)
  • Calibration routines
  • Display of all measured paramenters
  • Onboard Memory (Internal Data Logger)